Comments on: Comparative evaluation of sample digestion techniques for Atomic Absorption Spectroscopy https://lab-training.com/comparative-evaluation-of-sample-digestion-techniques-for-atomic-absorption-spectroscopy/ Experience Beyond Books Thu, 03 Sep 2020 13:14:32 +0000 hourly 1 https://wordpress.org/?v=6.7.2 By: Dr. Deepak Bhanot https://lab-training.com/comparative-evaluation-of-sample-digestion-techniques-for-atomic-absorption-spectroscopy/#comment-98392 Mon, 01 Oct 2018 05:07:39 +0000 https://lab-training.com/?p=30780#comment-98392 In reply to Duray.

You can use recommended methods if available for material specific analysis. Otherwise you can try acid digestion at room or elevated temperatures.Adoption of flame or graphite furnace atomization will depend on required concentration estimations. If results fall in ppm levels flame is recommended. However, for studies in ppb levels you should consider graphite furnace atomization.

]]>
By: Duray https://lab-training.com/comparative-evaluation-of-sample-digestion-techniques-for-atomic-absorption-spectroscopy/#comment-98312 Sun, 30 Sep 2018 07:59:23 +0000 https://lab-training.com/?p=30780#comment-98312 I want to convert plastic materials into homogenous liquids sample for AAS analysis.could you guide me in which digestion method is good for this purpose ? and which techniques flame or graphite AAS should I use ??? Please guide me .

]]>